[Handpiece Maintenance & Care]
1.Autoclave the handpiece before the first use, for 12 min. at 135℃.
2.after each treatment, sterilizing handpiece in autoclave according to the following steps:
1) Remove the bur and brush clean handpiece, clean outside of handpiece with an alcohol solution.
2) Lubricate handpiece, put lubrication oil into the air input holes in the connector.(recommend:dental handpiece lubrication oil).
3) Place into autoclaving pouch and mark.
4) Autoclave the handpiece for 12 min. at 135℃.
5) The tools which including spanner and descaling needle for changing the handpiece and working parts can be autoclaved for 12 min at 135℃.
3.Please send handpiece to the manufacturer or specified service agent to maintain when handpiece broke( not recommend field-repairable) .
4. Dental handpieces should not be soaked with acid disinfectant, otherwise the bearing and internal components of the phone will be seriously damaged.
5. Do not use an ultrasonic cleaner to clean the dental handpiece.